How should I treat a patient with an entrapped infected permanent pacemaker lead?
Autores de IdISBa
Autores ajenos a IdISBa
- Asmarats L
- Maristany J
- Pons J
- Buendía S
- Macaya F
- Pascual M
- Van Malderen S
- Szili-Torok T
- Lever N
- Webster M
Grupos de Investigación
Abstract
BACKGROUND: A 72-year-old man with an infected permanent dual chamber pacemaker. INVESTIGATION: The patient underwent successful generator removal, although the pacemaker leads were left due to severe entrapment despite energetic external traction, Percutaneous lead extraction was performed and was complicated by tricuspid valve avulsion leading to severe tricuspid regurgitation. DIAGNOSIS: Entrapment of infected pacemaker lead. MANAGEMENT: First, a loop around the leads was created using the combination of a gooseneck snare and a wire. Second, a single goose snare was used to remove the remaining severel gentrapped piece of lead.