A Randomized Comparison of Drug-Eluting Balloon Versus Everolimus-Eluting Stent in Patients With Bare-Metal Stent-In-Stent Restenosis

Fecha de publicación: Fecha Ahead of Print:

Autores de IdISBa

Autores ajenos a IdISBa

  • Alfonso F
  • Pérez-Vizcayno MJ
  • Cárdenas A
  • García Del Blanco B
  • Seidelberger B
  • Iñiguez A
  • Gómez-Recio M
  • Masotti M
  • Velázquez MT
  • Sanchís J
  • García-Touchard A
  • Zueco J
  • Melgares R
  • Cequier A
  • Dominguez A
  • Mainar V
  • López-Mínguez JR
  • Moreu J
  • Martí V
  • Moreno R
  • Jiménez-Quevedo P
  • Gonzalo N
  • Macaya C
  • RIBS V Study Investigators, under the auspices of the Working Group on Intervent

Grupos de Investigación

Abstract

Objectives This study sought to compare the efficacy of drug-eluting balloons (DEB) with that of everolimus-eluting stents (EES) in patients with bare-metal stents (BMS) in-stent restenosis (ISR). Background Treatment of patients with ISR remains a challenge. Methods This was a prospective, multicenter, randomized trial comparing DEB with EES in patients with bare-metal stents (BMS) in-stent restenosis (ISR). The primary endpoint was the minimal lumen diameter at 9 months' follow-up. Results A total of 189 patients with BMS-ISR from 25 Spanish sites were included (95 were allocated to DEB and 94 to EES). Procedural success was achieved in all patients. At late angiography (median 249 days; 92% of eligible patients), patients in the EES arm had a significantly larger minimal lumen diameter (2.36 +/- 0.6 mm vs. 2.01 +/- 0.6 mm, p < 0.001; absolute mean difference: 0.35 mm; 95% confidence interval [CI]: 0.16 to 0.53) and a lower percent of diameter stenosis (13 +/- 17% vs. 25 +/- 20%, p < 0.001). However, late loss (0.04 +/- 0.5 mm vs. 0.14 +/- 0.5 mm, p = 0.14) and binary restenosis rate (4.7% vs. 9.5%, p = 0.22) were very low and similar in both groups. Clinical follow-up (median 365 days) was obtained in all (100%) patients. Occurrences of the combined clinical outcome measure (cardiac death, myocardial infarction, and target vessel revascularization; 6% vs. 8%; hazard ratio [HR]: 0.76; 95% CI: 0.26 to 2.18, p = 0.6) and the need for target vessel revascularization (2% vs. 6%; HR: 0.32: 95% CI: 0.07 to 1.59, p = 0.17) were similar in the 2 groups. Conclusions In patients with BMS-ISR, both DEB and EES provided excellent clinical results with a very low rate of clinical and angiographic recurrences. However, compared with DEB, EES provide superior late angiographic findings. (Restenosis Intra-stent of Bare Metal Stents: Paclitaxel-eluting Balloon vs. Everolimus-eluting Stent [RIBS V]; NCT01239953) (C) 2014 by the American College of Cardiology Foundation

Datos de la publicación

ISSN/ISSNe:
0735-1097, 1558-3597

JACC-JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY  ELSEVIER SCIENCE INC

Tipo:
Article
Páginas:
1378-1386
PubMed:
24412457

Citas Recibidas en Web of Science: 226

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • device thrombosis; drug-eluting balloon; everolimus-eluting stent; in-stent restenosis