Diagnostic Performance of a Type III Portable Monitoring Device for Obstructive Sleep Apnea in Pregnant Women: A Prospective Validation Study.

Fecha de publicación: Fecha Ahead of Print:

Autores de IdISBa

Autores ajenos a IdISBa

  • Romero-Peralta S
  • Alonso M
  • Mediano O
  • Álvarez Ruiz De Larrinaga A
  • García-Río F

Grupos de Investigación

Abstract

Rationale: Obstructive sleep apnea (OSA) during pregnancy is linked to negative maternal and neonatal outcomes. Diagnosing OSA in this population is particularly challenging. Portable monitoring devices (PMD) present a potential alternative to polysomnography (PSG), but their effectiveness in pregnant women is uncertain. Objective: To evaluate diagnostic accuracy of a PMD to detect OSA in pregnant women. Methods: A prospective study of 136 middle-aged pregnant women (34 ± 4 years) in their third trimester of pregnancy (34 ± 3 weeks) who underwent laboratory type III PMD and PSG simultaneously. Results: Using an apnea-hypopnea index (AHI) =5 event/h by PSG, OSA was diagnosed in 10.3% of women (mild OSA: 86%; moderate OSA: 14%). An acceptable positive correlation was found between both tests in the AHI ( r = 0.787; p < 0.001) and oxygen desaturation index (ODI) ( r = 0.806; p < 0.001). The agreement limits between PSG and type III PMD were -4.1 to 5.4 for AHI and -6.0 to 4.5 for ODI. The sensitivity and specificity of type III PMD for an AHI =5 events/h were 57.1% and 99.2%, respectively, with a positive predictive value (PPV) of 88.9% and a negative predictive value (NPV) of 95.3%. On a receiver operating characteristic curve, the best cutoff point of AHI by type III PMD to identify OSA according to PSG criteria was 2.25 events/h. The sensibility and specificity for this point were 85.7% and 88.5%, respectively, with a PPV of 46.2% and NPV of 98.2%. Conclusions: A type III PMD with an AHI cutoff of 5 events/h could be a good alternative for OSA diagnosis in pregnant women. Additionally, an AHI =2.25 event/h demonstrates good diagnostic performance, but its low positive predictive value suggests that it is more appropriate as a screening tool. Further studies are needed to validate this sleep study tool in pregnant women, particularly in home settings.

Datos de la publicación

ISSN/ISSNe:
1540-9996, 1931-843X

JOURNAL OF WOMENS HEALTH  MARY ANN LIEBERT, INC

Tipo:
Article
Páginas:
685-693
PubMed:
39964806

Citas Recibidas en Web of Science: 2

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • diagnosis, obstructive sleep apnea, polysomnography, portable monitoring devices, pregnancy, validation study

Cita

Compartir