Early outcomes of the novel Myval THV series compared to SAPIEN THV series and Evolut THV series in individuals with severe aortic stenosis

Fecha de publicación:

Autores de IdISBa

Autores ajenos a IdISBa

  • Royen NV
  • Amat-Santos IJ
  • Hudec M
  • Bunc M
  • Ijsselmuiden A
  • Laanmets P
  • Unic D
  • Merkely B
  • Hermanides RS
  • Ninios V
  • Protasiewicz M
  • Rensing BJWM
  • Martin PL
  • Feres F
  • Sousa M
  • Belle EV
  • Linke A
  • Ielasi A
  • Montorfano M
  • Webster M
  • Toutouzas K
  • Teiger E
  • Bedogni F
  • Voskuil M
  • Pan M
  • Angerås O
  • Kim WK
  • Rothe J
  • Kristic I
  • Van den Branden BJL
  • Westermann D
  • Bellini B
  • Garcia-Gomez M
  • Tobe A
  • Tsai TY
  • Garg S
  • Thakkar A
  • Chandra U
  • Morice MC
  • Soliman O
  • Onuma Y
  • Serruys PW
  • Baumbach A

Grupos de Investigación

Abstract

BACKGROUND: There are limited head-to-head randomised trials comparing the performance of different transcatheter heart valves (THVs). AIMS: We aimed to evaluate the non-inferiority of the balloon-expandable Myval THV series compared to the balloon-expandable SAPIEN THV series or the self-expanding Evolut THV series. METHODS: The LANDMARK trial randomised 768 patients in a 1:1 ratio, (Myval THV series [n=384] vs contemporary series with 50% SAPIEN THV series [n=192] and 50% Evolut THV series [n=192]). The non- inferiority of Myval over the SAPIEN or Evolut THV series in terms of the 30-day primary composite safety and effectiveness endpoint as per the third Valve Academic Research Consortium (VARC-3) was tested in an intention- to-treat population with a predefined statistical power of 80% (1-sided alpha of 5%) for a non-inferiority margin of 10.44%. RESULTS: The Myval THV series achieved non-inferiority for the primary composite endpoint over the SAPIEN THV series (24.7% vs 24.1%, risk difference [95% confidence interval {CI}]: 0.6% [not applicable {NA} to 8.0]; p=0.0033) and the Evolut THV series (24.7% vs 30.0%, risk difference [95% CI]: -5.3% [NA to 2.5]; p<0.0001). The incidences of pacemaker implantation were comparable (Myval THV series: 15.0%, SAPIEN THV series: 17.3%, Evolut THV series: 16.8%). At 30 days, the mean pressure gradient and effective orifice area were significantly better with the Myval THV series compared to the SAPIEN THV series (p<0.0001) and better with the Evolut THV series than with the Myval THV series (p<0.0001). At 30 days, the proportion of moderate to severe prosthetic valve regurgitation was numerically higher with the Evolut THV series compared to the Myval THV series (7.4% vs 3.4%; p=0.06), while not significantly different between the Myval THV series and the SAPIEN THV series (3.4% vs 1.6%; p=0.32). CONCLUSIONS: The Myval THV series is non-inferior to the SAPIEN THV series and the Evolut THV series in terms of the primary composite endpoint at 30 days. Clinical trial registration: ClinicalTrials.gov: NCT04275726; EudraCT number 2020-000,137-40.

Datos de la publicación

ISSN/ISSNe:
1774-024X, 1969-6213

EuroIntervention  EUROPA EDITION

Tipo:
Article
Páginas:
105-118
PubMed:
39589296

Citas Recibidas en Web of Science: 14

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • aortic stenosis; balloon-expandable valve; non-inferiority; randomised trial; self-expanding valve; transcatheter heart valve