An HPLC-UV Method to Assess Human Plasma 25(OH)D3
Fecha de publicación:
Autores de IdISBa
Autores ajenos a IdISBa
- Tijerina A
- Garza A
- López A
- Cavazos N
- Romo A
- Heya MS
- Salas R
Grupos de Investigación
Abstract
The aim of this study was to validate an HPLC-UV method to assess vitamin D status by determining the linearity and precision of the 25-hydroxyvitamin D-3 (25(OH)D-3) calibration curve, the limits of detection, quantitation and robustness of the method, and its accuracy. A second stock solution of 25(OH)D-3 was prepared (500 ng/mL), and working dilutions (5, 10, 20, 30, 40, and 50 ng/mL) were prepared for a calibration curve. The HPLC equipment had a UV-Vis diode-array detector and utilized an Acclaim (TM) 120 C18 column (5 mu m, 4.6 x 250 mm) with a flow rate of 1.2 mL/min, a column temperature of 30 degrees C, and the standards and samples were maintained at 4 degrees C, with an injection volume of 100 mu L. Detection of 25(OH)D-3 was determined at 265 nm, with a retention time of 4.0 min. The validation was conducted according to the FDA Validation of Analytical Procedures: Guidance for Industry. Vitamin D was extracted from plasma samples using acetonitrile (ACN)-0.1% formic acid (2:1 v/v), and the percentage of recovery was calculated. The proposed method conditions gave excellent linearity (R-2 = 0.9989) and the linearity coefficient was R-2 > 0.99 for 25(OH)D-3. The detection and quantification limits were 1.1703 ng/mL and 3.5462 ng/mL, respectively. Decreasing or increasing the reading temperature by 1 degrees C decreased the response units (AU) of vitamin D, 25(OH)D-3. When the current flow rate decreased by 0.2 mL/min (1.0 mL/min), the retention time increased to 4.913 min, whereas an increase of 0.2 mL/min of the proposed flow rate (1.4 mL/min) decreased the retention time to 3.500 min. The percentage of recovery varied from 92.2% to 97.1%. The proposed method to quantify a vitamin D metabolite (25(OH)D-3) in human plasma samples was reliable and validated.
Datos de la publicación
- ISSN/ISSNe:
- 2072-6643, 2072-6643
- Tipo:
- Article
- Páginas:
- -
- DOI:
- 10.3390/nu16142304
- PubMed:
- 39064747
Nutrients MDPI AG
Citas Recibidas en Web of Science: 6
Documentos
- No hay documentos
Filiaciones
Keywords
- 25-hydroxyvitamin D-3; HPLC-UV; linearity; precision; robustness; accuracy
Campos de Estudio
Cita
Tijerina A; Garza A; López A; Cavazos N; Romo A; Heya MS; Bouzas C; Tur JA; Salas R. An HPLC-UV Method to Assess Human Plasma 25(OH)D<sub>3</sub>. Nutrients. 2024. 16 (14):