An MRI comparative image evaluation under diagnostic and radiotherapy planning set-ups using a carbon fibre tabletop for pelvic radiotherapy
Autores de IdISBa
Autores ajenos a IdISBa
- Sabater, S
- Pastor-Juan, MR
- Berenguer, R
- Lozano-Setien, E
- Andres, I
- Tercero-Azorin, MI
- Sevillano, M
- Rovirosa, A
- Korte, E
- Arenas, M
Abstract
Purpose.-To quantitatively and qualitatively evaluate and compare MR images acquired from the usual diagnostic set-up with images acquired in a radiotherapy set-up. Materials and methods.-MR images of a phantom and 18 prostate cancer patients imaging using the usual diagnostic set-up with a curved tabletop (DX-set-up) and using a radiotherapy set-up (RT-set-up) with a flat tabletop, were analysed retrospectively. Quantitative assessments were made by measuring signal-to-noise ratio (SNR) and contrast to noise ratio (CNR). Non-parametric tests were used. Qualitative assessments were made independently by three radiologists. Inter-rater reliability was measured as the percentage of agreement and Gwet's AC2 test. Results.-The use of the RT-set-up was linked to 19.3% SNR drop on the phantom image. Up to 24% SNR prostate drop was observed in patients' images. Up to 34% CNR drop was observed in patients' images. Although image quality of the RT-set-up was acceptable, all measured domains were worse than the corresponding DX-set-up. Additionally diagnostic images were linked to a better percentage of agreement among raters. Conclusions.-SNR, CNR and image quality were significantly worse when using the RT-set-up than with the DX-set-up although Image quality in the RT-set-up was acceptable. (C) 2019 Societe francaise de radiotherapie oncologique (SFRO). Published by Elsevier Masson SAS. All rights reserved.
Datos de la publicación
- ISSN/ISSNe:
- 1278-3218, 1769-6658
- Tipo:
- Article
- Páginas:
- 296-303
- PubMed:
- 31133514
Cancer Radiotherapie Elsevier Masson
Citas Recibidas en Web of Science: 2
Documentos
- No hay documentos
Filiaciones
Keywords
- MRI; SNR; CNR; Carbon fibre; Tabletop; Radiotherapy
Cita
Sabater S; Pastor Juan MR; Berenguer R; Lozano Setien E; Andres I; Tercero Azorin MI; Sevillano M; Jimenez E; Rovirosa A; Korte E; Arenas M. An MRI comparative image evaluation under diagnostic and radiotherapy planning set-ups using a carbon fibre tabletop for pelvic radiotherapy. Cancer Radiotherapie. 2019. 23 (4): 296-303.
Portal de investigación